This file describes the principle of AEQ-TR HIV, Sífilis e HCV, the target audience, the provided materials, and the required materials.
This instruction manual describes the AEQ-TR, an external quality assessment for rapid tests (TR) in which registered institutions receive, in each round, panels with four dried biological samples. The samples are hydrated and tested as if they were whole blood from four routine patients, using the HIV, syphilis and HCV rapid test kits supplied by the Ministry of Health. Any professional who performs rapid tests in services of the Ministry of Health network may take part.
Each panel holds enough sample volume for two participants and contains an instruction manual, two draft forms, four numbered tubes with 20µL to 30µL of dried samples, a tube with 1.8mL of hydration solution (PBS/Tween 20), a Pasteur pipette and a silica sachet. The manual covers storage and stability of the samples, the hydration procedure, testing, biosafety and references, and notes that results are entered individually and confidentially on the Quali-TR website. This is the 6th edition, revised in May 2018. This manual is written in Portuguese.
| Samples per panel | Four dried biological samples (tubes numbered 1 to 4) |
|---|---|
| Sample volume per tube | 20µL to 30µL of dried biological sample |
| Hydration solution supplied | 1.8mL of PBS/Tween 20 solution |
| Hydration solution added per tube | 7 drops (270µL) |
| Storage temperature of dried samples | 2°C to 30°C until the expiry date |
| Stability after hydration | 24 hours at 2°C to 30°C |
| Rest time after hydration | At least 2h and at most 24h |
Each panel contains four tubes numbered 1 to 4 with dried biological samples of unknown reactivity for HIV, syphilis and HCV.
Only Professional 1 hydrates the samples; Professional 2 then performs steps 5 to 7 and each professional tests all four samples individually.
Keep the panel with the dried samples between 2°C and 30°C until the expiry date on the back of the plastic packaging.
Each professional transcribes the data from the result forms into Quali-TR, under the section "Inserir Resultados", individually and confidentially.