This file is a product manual for CANopen® slave interface for digital multibus Mass Flow / Pressure instruments. It introduces the function, feature and usage method of the product.
This instruction manual describes the CANopen slave interface for Bronkhorst digital multibus Mass Flow / Pressure meters and controllers. The interface provides a direct connection to CANopen networks and supports the CiA 404 device profile for measuring devices and closed-loop controllers. The manual covers installation of the CANopen network, instrument configuration with the EDS file, PDO and SDO setup, the object dictionary, and troubleshooting.
Chapters explain the M12-A coded connector and cabling, network termination, Node-ID and data rate settings via rotary switches or software, the NMT state machine and heartbeat, and LED indications for diagnostics. Appendices list the object dictionary and SI and non-SI unit specifications.
| Connector | Micro-style sealed M12-A coded connector |
|---|---|
| PDOs | 4 RPDOs and 4 TPDOs with 8 bytes of process data each |
| Network termination | 121 Ohms, 1%, 1/4W terminating resistors at both ends |
| Device profile | CiA 404 device profile for measuring devices and closed-loop controllers |
Load the Bronkhorst CANopen EDS file named Bronkhorst_Meter_Controller_CANopen.EDS into the PLC software, then add the Bronkhorst Meter Controller device to the CANopen master.
The CAN trunk line must be terminated at both ends with 121 Ohms, 1%, 1/4W terminating resistors.
Bronkhorst products are warranted against defects in material and workmanship for three years from the date of shipment, provided they are used according to the ordering specifications and this manual.
Bit 0 of object 0x6425:01 "CO Control Byte" should be 1 to activate the PID controller; it is activated by default but can be overwritten by the master via process data.