The Norgen's Magnetic Bead Stool DNA Isolation Kit, which can be used on the Hamilton Vantage Liquid Handler, has been compared to a manual method using magnetic beads and a competitor's manual method using a spin column. The results indicate that the automated system yielded 6X more DNA (~ 200 ng/μL) than the manual methods, and that the DNA derived from the automated method was of high purity.
This Application Note 102 from Norgen Biotek Corp. compares DNA isolated from stool using Norgen's Stool DNA Isolation 96-Well Kit (Magnetic Bead System), Cat. 63100, run on an automated Hamilton Vantage Liquid Handler, with Norgen's manual magnetic bead method (Cat. 55700) and a competitor's manual spin column method. Preserved stool was collected in Norgen's Stool Nucleic Acid Collection and Preservation Tubes (Cat. 45660) and aliquoted into 48 samples of 200 μL, 16 per method.
The automated method, which uses a bead beating lysis step followed by magnetic bead binding, an optimized wash protocol and elution, gave 260 ng/μL compared with 33 ng/μL for the manual methods, with good 260/280 and 260/230 purity ratios. qPCR with increasing eluate volumes (2, 4, 8 μL) showed negligible PCR inhibition, and 16S sequencing with the 16S V3-V4 Library Preparation Kit for Illumina (Cat. 70400) showed good agreement in community composition between manual and automated samples.
| Automated platform used | Hamilton Vantage Liquid Handler |
|---|---|
| Automated kit catalog number | 63100 |
| Automated kit size | 2 x 96-Well Plates |
| Manual kit catalog number | 55700 |
| Sample aliquot volume | 200 μL |
| DNA yield, automated method | 260 ng/μL |
| Gel electrophoresis conditions | 1.2% agarose gel at 170 V for 25 minutes |
The automated Norgen method on the Hamilton Vantage gave 260 ng/μL, compared with 33 ng/μL for the manual methods.
Norgen's Stool DNA Isolation 96-Well Kit (Magnetic Bead System) was optimized for use by the Hamilton Vantage Liquid Handler.
No; qPCR of increasing eluate volumes from 2 to 8 μL showed negligible PCR inhibitors in the samples.
Permutational ANOVA did not show any significant difference between the manual and automated samples.