DeVilbiss DV63 Standard Plus and DV64 AutoPlus CPAP Series Provider Quick Setup Guide

This document introduces the quick setup guide of DeVilbiss DV63 standard and DV64 auto CPAP series devices, including device settings, entering clinical menu, turning on/off options, etc.


DeVilbiss DV63 Standard Plus and DV64 AutoPlus CPAP Series Provider Quick Setup Guide - cover page
Brand
DeVilbiss
Category
Home medical equipment
Document type
Quick Start Guide
Language
English
Pages
12
File format
PDF
File size
1.6 MB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
F9CBACB4369D621180835B3FADCC5FEA

About this manual

Provider Quick Setup Guide for the DeVilbiss DV63 Standard Plus and DV64 AutoPlus CPAP series. It explains how qualified medical professionals access the clinical menu and adjust device settings; it states that these setup instructions should not be provided to the patient.

The guide covers main menu items, language, Mask Fit Check, Quick View, CPAP mode and pressure settings, SmartFlex, ramp time and pressure, humidity, tubing, volume, brightness, usage threshold, enable/show/hide/lockout options, algorithm, admin mode, clock settings, and add-ons.

What's inside

  • Accessing Device Settings
  • Main Menu Items
  • Language Screen
  • Mask Fit Check Screen
  • Quick View Screen
  • CPAP Settings Screen - Mode
  • CPAP Settings - Pressure Settings
  • CPAP Settings - SmartFlex Settings
  • CPAP Settings - Ramp Time & Pressure Settings
  • Clock Settings

Specifications

Pressure adjustment increment0.5
Pressure unit optionscmH2O, hPa, mBar
Ramp time range0-45 minutes in 5 minute increments
Tubing diameter options15 mm, 22 mm
Tubing length options1.8 m (6'), 2.4 m (8'), 3.0 m (10'), 3.7 m (12')
Usage threshold range3 - 9 hours
Maximum humidity setting5 bars
DV64 default modeAuto
CPAP pressure initial/default setting8 / 10

Frequently asked questions

How do I access the clinical menu on a DV63 or DV64 device?

With the device plugged in, press the POWER button, RAMP button, and ROTARY KNOB at the same time. The display shows “Entering Clinical Menu.”

How long does the device remain in the Clinical Menu without user interaction?

It remains in the Clinical Menu for 15 minutes without user interaction before automatically exiting.

What pressure unit options are available?

The pressure unit can be toggled between cmH2O, hPa, and mBar.

Can the patient change the mode?

No, the mode cannot be changed by the patient.