This document introduces the quick setup guide of DeVilbiss DV63 standard and DV64 auto CPAP series devices, including device settings, entering clinical menu, turning on/off options, etc.
Provider Quick Setup Guide for the DeVilbiss DV63 Standard Plus and DV64 AutoPlus CPAP series. It explains how qualified medical professionals access the clinical menu and adjust device settings; it states that these setup instructions should not be provided to the patient.
The guide covers main menu items, language, Mask Fit Check, Quick View, CPAP mode and pressure settings, SmartFlex, ramp time and pressure, humidity, tubing, volume, brightness, usage threshold, enable/show/hide/lockout options, algorithm, admin mode, clock settings, and add-ons.
| Pressure adjustment increment | 0.5 |
|---|---|
| Pressure unit options | cmH2O, hPa, mBar |
| Ramp time range | 0-45 minutes in 5 minute increments |
| Tubing diameter options | 15 mm, 22 mm |
| Tubing length options | 1.8 m (6'), 2.4 m (8'), 3.0 m (10'), 3.7 m (12') |
| Usage threshold range | 3 - 9 hours |
| Maximum humidity setting | 5 bars |
| DV64 default mode | Auto |
| CPAP pressure initial/default setting | 8 / 10 |
With the device plugged in, press the POWER button, RAMP button, and ROTARY KNOB at the same time. The display shows “Entering Clinical Menu.”
It remains in the Clinical Menu for 15 minutes without user interaction before automatically exiting.
The pressure unit can be toggled between cmH2O, hPa, and mBar.
No, the mode cannot be changed by the patient.