Carbolite TZF & GZF 1200°C 3-Zone Tube Furnace Installation, Operation & Maintenance Instructions

This is an instruction manual that introduces the installation, operation and maintenance instructions for the 1200°C tube furnaces (3 zones) models TZF and GZF.


Carbolite TZF & GZF 1200°C 3-Zone Tube Furnace Installation, Operation & Maintenance Instructions - cover page
Brand
Carbolite
Category
Heater
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
137 KB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
31C717E981FDAD9E49F6B0E7DD357A3E

About this manual

This manual covers the Carbolite TZF and GZF three-zone tube furnaces rated for 1200°C, which are designed to achieve a single heating zone of extended temperature uniformity. The TZF range uses a ceramic work tube wound with resistance wire, while the GZF range uses modular heating elements of resistance wire partially embedded in lightweight refractory fibre. Controller manuals are supplied separately and should be read before operating the furnace.

The document explains unpacking and handling, siting and electrical connections, the three available 3-zone control methods (back-to-back thermocouples, broadcast comms retransmission, and independent zones), plus maintenance, repairs, fault analysis, circuit details, fuses and power settings, and specifications.

What's inside

  • Introduction
  • Installation
  • Operation
  • Maintenance
  • Repairs & Replacements
  • Fault Analysis
  • Circuit Details
  • Fuses & Power Settings
  • Specifications

Specifications

Maximum temperature1200°C
Element module length182mm
End element rating1500W maximum
Centre element rating1000W

Frequently asked questions

How should I lift or move the furnace?

Always lift it by its base or by both ends of the main body, never by the work tube or surrounding insulation; for larger models use two people.

Can GZF 3-phase models be rewired for single-phase supply?

Yes, from Year 2000 GZF models ordered for 3-phase supply can be rewired between 1-phase and 3-phase supplies in the ranges 380/220V–415/240V and 208-240V, but not 3-phase star without neutral.

What are the three 3-zone control methods?

Method A uses back-to-back thermocouples, method B retransmits the setpoint using Eurotherm 3000 series broadcast comms, and method C uses independent control zones.

What should I do if a furnace with back-to-back thermocouple control does not cool down?

The end controllers may be trying to remain at the central temperature; contact Carbolite's technical department.

Download PDF (137 KB)

More Carbolite Manuals


View all Carbolite manuals →