Draeger Aerotest Compressed Air Quality Meter Oil and Water Vapour Test Tube Instruction Manual

This manual describes the usage and principles of the oil test tube of the Draeger compressed air quality meter.


Draeger Aerotest Compressed Air Quality Meter Oil and Water Vapour Test Tube Instruction Manual - cover page
Brand
Draeger
Category
Other products
Document type
User Manual
Language
Chinese
Pages
11
File format
PDF
File size
3.1 MB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
7642C67FBF6945ADD6FF19DABCEADC27

About this manual

This document compiles Chinese-language instructions for several Draeger test tubes used with the Aerotest compressed air quality meter. It covers the Impactor oil detection box, Oil 10/a-P oil test tube, Water vapour 20/a-P test tube, and Water Vapour 5/a-P test tube.

It describes application ranges, measurement principles, operating steps, evaluation tables, cross sensitivity, storage, and disposal notes for these test tubes. This manual is written in Chinese.

What's inside

  • Impactor oil detection box
  • Oil 10/a-P test tube
  • Water vapour 20/a-P test tube
  • Water Vapour 5/a-P test tube
  • Measurement and evaluation
  • Cross sensitivity
  • Additional information
  • Preconditions

Specifications

Impactor oil detection box measurement range0.1, 0.5, or 1.0 mg/m3
Impactor oil detection box detection limit0.05 mg/m3
Impactor oil detection box temperature10 to 30 C
Impactor oil detection box humidityMaximum 60% relative humidity
Impactor oil detection box pressureOnly for unpressurized compressed air
Oil 10/a-P measurement range0.1, 0.3, 0.5, or 1.0 mg/m3
Oil 10/a-P color changeWhite to light brown or yellow
Oil 10/a-P temperature10 to 30 C
Oil 10/a-P pressureOnly for depressurized compressed air
Water vapour 20/a-P measurement range25 to 250, 35 to 500, or 150 to 1500 mg/m3
Water vapour 20/a-P measurement time10, 5, or 2.5 minutes
Water Vapour 5/a-P flow rate2 L/min
Water Vapour 5/a-P measurement range at 25 L sample10 to 450 mg H2O/m3

Frequently asked questions

What is the Impactor oil detection box used for?

It is used to detect oil mist in compressed air.

Which meters can be used with Oil 10/a-P test tubes?

They are used only with Draeger Aerotest D20700, Aerotest Light, and Aerotest Simultan.

What is the measurement range of the Water vapour 20/a-P test tube?

The measurement range is 25 to 250, 35 to 500, or 150 to 1500 mg/m3.

What flow rate is required for the Water Vapour 5/a-P test tube?

The flow rate must be set to 2 L/min.