This manual describes the usage and principles of the oil test tube of the Draeger compressed air quality meter.
This document compiles Chinese-language instructions for several Draeger test tubes used with the Aerotest compressed air quality meter. It covers the Impactor oil detection box, Oil 10/a-P oil test tube, Water vapour 20/a-P test tube, and Water Vapour 5/a-P test tube.
It describes application ranges, measurement principles, operating steps, evaluation tables, cross sensitivity, storage, and disposal notes for these test tubes. This manual is written in Chinese.
| Impactor oil detection box measurement range | 0.1, 0.5, or 1.0 mg/m3 |
|---|---|
| Impactor oil detection box detection limit | 0.05 mg/m3 |
| Impactor oil detection box temperature | 10 to 30 C |
| Impactor oil detection box humidity | Maximum 60% relative humidity |
| Impactor oil detection box pressure | Only for unpressurized compressed air |
| Oil 10/a-P measurement range | 0.1, 0.3, 0.5, or 1.0 mg/m3 |
| Oil 10/a-P color change | White to light brown or yellow |
| Oil 10/a-P temperature | 10 to 30 C |
| Oil 10/a-P pressure | Only for depressurized compressed air |
| Water vapour 20/a-P measurement range | 25 to 250, 35 to 500, or 150 to 1500 mg/m3 |
| Water vapour 20/a-P measurement time | 10, 5, or 2.5 minutes |
| Water Vapour 5/a-P flow rate | 2 L/min |
| Water Vapour 5/a-P measurement range at 25 L sample | 10 to 450 mg H2O/m3 |
It is used to detect oil mist in compressed air.
They are used only with Draeger Aerotest D20700, Aerotest Light, and Aerotest Simultan.
The measurement range is 25 to 250, 35 to 500, or 150 to 1500 mg/m3.
The flow rate must be set to 2 L/min.