The quick reference guide introduces the operation of Draeger Apollo ventilator.
This guide provides concise instructions for starting the Draeger Apollo ventilator, adjusting ventilation modes (volume, pressure, pressure support), and handling common alarms. It includes steps for self-testing, manual ventilation, and end-of-day procedures such as turning off the machine and checking vaporizers and soda lime. The guide clarifies that it is not a replacement for the full operator’s manual.
Volume Control sets a target tidal volume; Pressure Control limits peak inspiratory pressure and allows monitoring of tidal volume and minute ventilation.
Red indicates a warning (high priority), yellow indicates a caution (medium priority), and white indicates an advisory (low priority).