Top Instrument SPX Series Biochemical Incubator User Manual

This is a user manual of SPX series smart biochemical incubator.


Top Instrument SPX Series Biochemical Incubator User Manual - cover page
Brand
Top Instrument
Category
Other
Document type
User Manual
Language
Chinese
Pages
16
File format
PDF
File size
759 KB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
836216186D52CCF8ED016B8A2ED3117D

About this manual

This is the Chinese user manual for the SPX Series intelligent biochemical incubator made by Zhejiang Top Instrument Co., Ltd. (Top Instrument). The manual covers the models SPX-150 through SPX-1500, with volumes from 150 L to 1500 L, and describes a unit with both cooling and heating temperature control used for low-temperature constant-temperature tests, culture tests and environmental tests in laboratories.

The manual explains safety precautions, the standard items in the package, the main technical data, the cabinet and panel structure, and the key functions used to set programs and parameters. It also includes an electrical wiring diagram, maintenance and troubleshooting information, warranty terms and contact details. The cabinet is described as having a mirror-finish stainless steel inner chamber, casters, a glass viewing window and an independent over-temperature alarm system that interrupts operation when the limit temperature is exceeded. This manual is written in Chinese.

What's inside

  • Overview
  • Unpacking Check of Standard Accessories
  • Instrument Specifications
  • Main Technical Data
  • Composition and Structure (Cabinet and Panel)
  • Operation and Use (Keys and Settings)
  • Electrical Wiring Diagram
  • Maintenance and Diagnosis
  • Warranty Information
  • Contact Information

Specifications

Temperature setting range0 °C to 50 °C
Temperature fluctuation±0.5 °C
Temperature deviation±1.0 °C
NoiseLess than 70 dB
Heating power300 W
Cooling power150 W to 350 W
Power supplyAC 220 V, 50-60 Hz
Compressor operationIntermittent, 3-minute shutdown delay protection
Operating environment0-35 °C, humidity ≤85%RH, 86-106 kPa
Model volumes150 L (SPX-150) up to 1500 L (SPX-1500)

Frequently asked questions

What is the temperature control range of the SPX Series incubator?

The temperature setting range is 0 °C to 50 °C, with a temperature fluctuation of ±0.5 °C and a temperature deviation of ±1.0 °C.

What is included in the standard package?

The package contains one SPX Series incubator, one product manual, one certificate of conformity, one warranty card and one water collection box.

What should I do if abnormal odor or smoke appears during use?

Switch off the power immediately and do not attempt repairs yourself; contact the company's after-sales service so a professional technician can check the unit.

What power supply does the incubator require?

It uses AC 220 V, 50-60 Hz, and the socket must have a good ground connection.