AT817 is a precision Lcr digital bridge produced by Applent Technology. Its main features are basic accuracy of 0.05%, ten-point test frequency of 50 Hz ~ 100 kHz, test level of 0.1 V rms, 0.3 V rms, 1 V rms, output impedance of 30 Ω, 100 Ω, series and parallel equivalent, fast test of 15 times / second, built-in 20 group comparator records, 1-stage and 3-stage selection, built-in master-slave dual-mode trigger Handler interface, and built-in rs232c interface
This is the Simplified Chinese operation manual (Rev.A, Oct 2006) for the Applent AT817 precision LCR digital bridge, a compact bench instrument controlled by a microprocessor. It measures inductance L, capacitance C, resistance R, complex impedance Z, quality factor Q, dissipation factor D and phase angle θ, with stated accuracy up to 0.05%, and is intended for component makers, schools, research institutes and metrology/inspection departments.
The manual opens with safety notices, warranty terms and a table of contents, then walks through installation and setup (packing list, power requirements, fuse replacement, operating environment, cleaning), an overview of features and specifications, and getting started: front panel, rear panel, power-on, and test setup for frequency, level, source resistance, parameters, rate, equivalent mode, ranging, zeroing, remote control, trigger, admin settings and key lock. Later chapters describe the comparator function, the Handler interface wiring and timing, performance testing, and an appendix of specifications. This manual is written in Chinese.
| Measurement parameters | L, C, R, Z, D, Q, θ (degrees), θ (radians) |
|---|---|
| Accuracy | up to 0.05% |
| Test frequency | 50Hz, 100Hz, 120Hz, 1kHz, 10kHz, 20kHz, 40kHz, 50kHz, 100kHz |
| Frequency accuracy | ±0.02% |
| Test level | 0.1V, 0.3V, 1.0V (three levels), accuracy ±10% |
| Source resistance | 30Ω, 50Ω or 100Ω (constant voltage source) |
| Measurement speed | Fast, medium and slow; fastest 15 tests |
| Ranges | Six ranges, automatic or manual |
| Equivalent modes | Series and parallel |
| Interfaces | Handler interface and RS232C |
| Memory | 20 groups of stored data |
| Power requirement | 198-252VAC, 47.5-52.5Hz, max 15VA |
| Fuse | 250V, 0.5A slow blow |
| Operating environment | 0°C to 55°C; humidity below 95%RH at 40°C |
It automatically measures inductance L, capacitance C, resistance R, complex impedance Z, quality factor Q, dissipation factor D, and phase angle θ in both degrees and radians.
The manual states an accuracy of up to 0.05%.
It must be used with 198-252VAC, 47.5-52.5Hz supply and draws a maximum of 15VA.
It is equipped with a Handler interface and an RS232C interface, and its remote commands are compatible with SCPI.