BMG LABTECH POLARstar Omega Microplate Reader Europium Assay Setup Guide

The document mainly introduces the configuration method of Europium-based Assay on BMG LABTECH POLARstar Omega Microplate Reader.


BMG LABTECH POLARstar Omega Microplate Reader Europium Assay Setup Guide - cover page
Brand
BMG Labtech
Category
medical technology
Document type
Application Note
Language
English
Pages
17
File format
PDF
File size
1.5 MB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
C4C4DDE3ECDE96358630A01608D7F7FF

About this manual

This setup guide describes how the BMG LABTECH POLARstar Omega microplate reader is configured for Invitrogen's Europium-based TR-FRET assays. The reader was tested with the Adapta Universal Kinase Assay Kit (PV5099) and poly E4Y substrate against JAK2 JH1/JH2 and JAK2 JH1/JH2 V617F kinases, as well as the LanthaScreen Eu Kinase Binding Assay.

The document lists the recommended optics (337 nm excitation, 665-10 and 620-10 HTRF emission filters, optic module height setting 11) and then walks through the Omega Control software: creating a Time Resolved Fluorescence endpoint test protocol, entering integration start and time of 100 and 200 µs, setting two multichromatics, selecting filters and wells, and adjusting donor and acceptor gain. It closes with assay plate preparation steps, a plate schematic and representative results.

What's inside

  • Recommended Optics
  • Instrument Setup
  • Creating a Test Protocol
  • Time Resolved Fluorescence and Endpoint selection
  • Integration start and time settings
  • Filter selection and well layout
  • Gain adjustment for donor and acceptor channels
  • Assay plate preparation
  • Assay Plate Schematic
  • Results

Specifications

Excitation filter337 nm
HTRF Emission 1 filter665-10
HTRF Emission 2 filter620-10
Optic module height setting11
Integration settingsIntegration: Start 100 µs, Integration: Time 200 µs
Number of multichromatics2

Frequently asked questions

Which filters and height setting are recommended for Europium-based TR-FRET?

Use a 337 nm excitation filter with the 665-10 and 620-10 HTRF emission filters, and set the optic module height setting to 11.

What integration settings should be entered in the protocol?

Set Integration: Start to 100 µs and Integration: Time to 200 µs.

Which optic head is needed and how should its dial be set?

The TRF Optic Head from BMG is required, and on the TRF optic the dial should be set to 11 for this application.

How is gain adjusted before measuring the plate?

A low-FRET well is selected to adjust the donor gain in the 620 channel, then a high-FRET well is selected to optimize the acceptor gain in the 665 emission channel.