Key facts
- Manuals
- 1
- Languages
- EN
- Last added
- 2023-10-30
Featured BMG Labtech manuals

BMG LABTECH POLARstar Omega Microplate Reader Europium Assay Setup Guide
This setup guide describes how the BMG LABTECH POLARstar Omega microplate reader is configured for Invitrogen's Europium-based TR-FRET assays. The reader was tested with the Adapta Universal Kinase Assay Kit (PV5099) and poly E4Y substrate against JAK2 JH1/JH2 and JAK2 JH1/JH2 V617F kinases, as well as the LanthaScreen Eu Kinase Binding Assay.
The document lists the recommended optics (337 nm excitation, 665-10 and 620-10 HTRF emission filters, optic module height setting 11) and then walks through the Omega Control software: creating a Time Resolved Fluorescence endpoint test protocol, entering integration start and time of 100 and 200 µs, setting two multichromatics, selecting filters and wells, and adjusting donor and acceptor gain. It closes with assay plate preparation steps, a plate schematic and representative results.
Covers: Recommended Optics · Instrument Setup · Creating a Test Protocol · Time Resolved Fluorescence and Endpoint selection · Integration start and time settings · Filter selection and well layout
Latest BMG Labtech manuals
BMG LABTECH POLARstar Omega Microplate Reader Europium Assay Setup Guide
Application Note 17 pages · 1.5 MB · ENGuide to setting up Invitrogen Europium-based TR-FRET assays, including Adapta and LanthaScreen kinase assays, on the BMG LABTECH POLARstar Omega reader.
Which document do I need?
Application Note
How to use the product for one specific task, with worked examples.
Questions answered in BMG Labtech manuals
Which filters and height setting are recommended for Europium-based TR-FRET?
Use a 337 nm excitation filter with the 665-10 and 620-10 HTRF emission filters, and set the optic module height setting to 11.
From: BMG LABTECH POLARstar Omega Microplate Reader Europium Assay Setup Guide
What integration settings should be entered in the protocol?
Set Integration: Start to 100 µs and Integration: Time to 200 µs.
From: BMG LABTECH POLARstar Omega Microplate Reader Europium Assay Setup Guide