Introduced the characteristics, installation method and working principle of ECH2O EA-10 soil moisture sensor
This guide explains how to install and operate the Decagon EA-10 soil moisture sensor, including wiring, calibration, and troubleshooting. It covers installation methods for turf and general soil, electrical connections, and functional testing. The sensor uses a 4-20 mA analog output to measure volumetric water content, with specifications for accuracy, temperature range, and power requirements. Calibration equations are provided for converting current to water content, and troubleshooting tips address common errors like poor probe-soil contact.
| Model | ECH2O EA-10 Dielectric Soil Moisture Sensor |
|---|---|
| Interface | 4-20 mA, 2-wire analog transmitter |
| Supply Voltage | 7-32 Volt DC |
| Measurement Range | 0 to 40 percent VWC |
| Measurement Resolution | depends on acquisition hardware |
| Measurement Accuracy | ± 2% with soil specific calibration |
| Active Sensor Length | 10 cm |
| Minimum Measurement Time | 10 ms |
| Operating Temperature | -40 to +60 °C |
| Dimensions | 14.5cm x 3.17cm x 0.15 cm |
| Cable Length | 3 m |
| Calibration Equation (Percent VWC) | Percent VWC = 4.25 x Current - 34.2 |
| Sensor Output in Air | 4.3 - 4.9 mA |
| Sensor Output in Distilled Water | 17.6 - 18.2 mA |
Percent Volumetric Water Content = 4.25 x Current - 34.2, where current is in milliamps.
The EA-10 has a 30-day satisfaction guarantee and a one-year warranty.
The EA-10 requires 7-32 Volt DC, and is reverse-polarity protected.