Lianhua LH-MUP230 (V11S) Portable Multi-Parameter Water Quality Tester Quick Start Guide

LH-MUP230(V11S)type portable multi-parameter water quality tester quick user guide, introduces the use of the instrument, including the instrument mode switching, COD measurement quick start, measurement process diagram and measurement operation process


Lianhua LH-MUP230 (V11S) Portable Multi-Parameter Water Quality Tester Quick Start Guide - cover page
Brand
Lianhua
Category
Analyzer
Document type
Quick Start Guide
Language
Chinese
Pages
9
File format
PDF
File size
1.1 MB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
D9ECF7C6A5218CBD7559B2D0949ED6A8

About this manual

The LH-MUP230 (V11S) is a portable multi-parameter water quality tester from Lianhua. This quick start guide explains how to switch the LH-9AE portable digester between its preset COD, total phosphorus and total nitrogen modes and how to prepare the instrument for measurement.

For each parameter, the guide lists the recommended sample volume and curve, then walks through reagent addition, digestion or color development, cooling, blanking and colorimetric reading. Sections cover COD, ammonia nitrogen, total phosphorus, total nitrogen and turbidity. This manual is written in Chinese.

What's inside

  • Digester mode switching
  • COD measurement quick start
  • Ammonia nitrogen measurement quick start
  • Total phosphorus measurement quick start
  • Total nitrogen measurement quick start
  • Turbidity measurement quick start

Specifications

COD digestion temperature165 °C
COD digestion time10 min
Total phosphorus digestion temperature120 °C
Total phosphorus digestion time30 min
Total nitrogen digestion temperature122 °C
Total nitrogen digestion time40 min

Frequently asked questions

How do I enter the digestion mode selection on the LH-MUP230 (V11S)?

With the instrument off, hold the temperature key while turning the power on, keep holding for more than 2 seconds, then release to enter mode selection.

What does the display show after pressing the blank key?

The display shows “0.000mg/L” after the blank key is pressed.

How much water sample is used in the COD procedure?

The procedure uses 2.0 mL of water sample placed in an LH-COD-YZ51 reagent tube.

What are the preset digestion modes?

The instrument has preset modes for COD at 165 °C, total phosphorus at 120 °C, and total nitrogen at 122 °C.