Oxford Instruments CMI233 Coating Thickness Gauge User Operation Manual

CMI233 Coating Thickness Gauge is mainly used for non-destructive measurement of coating thickness. It can be used for non-magnetic coating layers on steel, such as paint, plastic, enamel, chrome, zinc, etc., and all insulating layers on non-ferrous metals (such as copper, aluminum, austenitic stainless steel), such as anodic oxide film, paint, coating, etc.


Oxford Instruments CMI233 Coating Thickness Gauge User Operation Manual - cover page
Brand
Oxford Instruments
Category
Test Set
Document type
User Manual
Language
Chinese
Pages
7
File format
PDF
File size
207 KB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
7C2E9AB851AAA5E3EF9D37415B2B69CC

About this manual

This Chinese-language user operation manual covers the Oxford Instruments CMI233 coating thickness gauge. It explains setup, cable connection, power on/off, probe use, unit switching, probe tuning, measurement modes, and two-point calibration.

The manual also describes statistics viewing and clearing, calibration storage, upper and lower limit settings, data download, and basic troubleshooting. This manual is written in Chinese.

What's inside

  • 仪器简介
  • 设置
  • 探头调谐
  • 测量
  • 两点校准
  • 统计功能
  • 校准存储
  • 上下限设置
  • 数据下载
  • 疑难解答

Specifications

Keyboard model16-key model
Data storage99 data storage groups
Magnetic induction calibration groups01–50
Eddy current calibration groups51–99
Auto power offAfter 10 seconds of no use

Frequently asked questions

How do I tune the probe on the CMI233?

Connect the probe and turn the unit on, then press the * key, enter 111, and press ENTER. The display shows 'tune' for about 5 seconds and the unit turns off.

How do I switch measurement units between mils and micrometers?

Press the ZERO/UNITS key to convert MILS to µM.

How do I clear the statistics on the CMI233?

Press MEAN to enter statistics mode, then press CLEAR; press CLEAR again to confirm and clear all data.

How do I reset the CMI233 to factory settings?

Turn the unit on and press *9191; this resets the instrument to factory settings and turns it off.