OptiQuantTM Sludge Interface Monitor is a device used to measure the thickness of the sludge layer and the height of the sludge interface. It uses an ultrasonic measurement method with a measurement range of 0.2-12 meters.
This manual provides detailed instructions for operating, installing, and maintaining the Hach OptiQuantTM Sludge Interface Monitor. It includes safety precautions, technical specifications, setup procedures, calibration, maintenance, and troubleshooting. The device uses ultrasonic technology to measure sludge interface depth or thickness, with features like automatic cleaning and graphical display. It is designed for industrial applications where accurate sludge level monitoring is required.
The manual also covers installation steps, probe mounting, electrical connections, alarm relay setup, and maintenance tasks such as cleaning the window and replacing the wiper. Safety warnings and hazard information are included to ensure safe operation and prevent damage to the equipment or injury to personnel. This manual is written in Chinese.
| Measurement Range | 0.2-12 meters (sludge interface depth or thickness) |
|---|---|
| Resolution | 0.03 meters |
| Measurement Accuracy | <0.1 meters |
| Measurement Interval | 10 to 600 seconds |
| Output Signal | 0-20mA or 4-20mA, max load 500 ohms |
| Interface | DIN fieldbus (standard) |
| Enclosure Rating | IP65, suitable for outdoor installation |
| Operating Temperature | Probe: +2°C to +40°C; Monitor: -10°C to +40°C |
| Power Requirements | 230V AC, 50Hz, 14VA (supplied by user) |
| Alarm Relays | One upper limit relay and one lower limit relay, SPDT, 24V@2A |
| Dimensions | Probe: 15.5cm x 11.3cm diameter; Monitor: 28.6cm x 30.5cm x 9.6cm |
| Weight | Probe: ~2.33 kg; Monitor: ~3.6 kg |
| Certifications | GS and CE marks |
The measurement range is 0.2-12 meters for sludge interface depth or thickness.
The output signal is 0-20mA or 4-20mA, with a maximum load of 500 ohms.
The monitor operates between -10°C to +40°C, while the probe operates between +2°C to +40°C.
The device has GS and CE certifications.