Sinomeasure U-YZDPTU300-LCCN1 Turbidity Online Analyzer User Manual

This turbidity online analyzer is a patented product developed specifically for drinking water quality online monitoring. It has the characteristics of ultra-low turbidity detection limit, high precision measurement, long-term maintenance-free equipment, water-saving work and digital output, supporting cloud platform and mobile terminal data remote monitoring, as well as RS485-modbus communication. It can be widely used in tap water factory water, secondary water supply, pipeline end water, drinking water, membrane filtered water, swimming pool, surface water and other turbidity online monitoring.


Sinomeasure U-YZDPTU300-LCCN1 Turbidity Online Analyzer User Manual - cover page
Brand
Sinomeasure
Document type
User Manual
Language
Chinese
Pages
39
File format
PDF
File size
1010 KB
Published
29 October, 2023
Updated
29 October, 2023
MD5 checksum
FA27FFEA8A171176A716607FFAEEA235

About this manual

This user manual provides detailed instructions for operating and maintaining the Sinomeasure U-YZDPTU300-LCCN1 turbidity online analyzer. It covers safety information, technical specifications, mechanical installation, operational procedures, calibration methods, and data interfaces including Modbus RTU and 4-20mA. The device uses laser scattering to measure water turbidity and supports remote monitoring via cloud and mobile platforms.

The analyzer is designed for continuous monitoring of water turbidity in applications such as drinking water, swimming pools, and surface water. It features IP54 protection, a 24V DC power supply, and supports multiple measurement ranges (0-1NTU, 0-20NTU, 0-100NTU). Maintenance tasks include checking water flow, sensor condition, and cleaning the chamber. This manual is written in Chinese.

What's inside

  • Safety Information
  • Product Overview
  • Working Principle
  • Technical Specifications
  • Installation
  • Operation
  • Calibration
  • Maintenance
  • Data Interfaces (Modbus RTU and 4-20mA)

Specifications

Detection MethodLaser 90° scattering
Main StandardJJG880-2006
MaterialComposite material
Dimensions183mm*349mm*113.5mm (W*H*D)
Weight4.5Kg
Power SupplyDC24V
Protection RatingIP54
Installation TypeWall-mounted
Operating Temperature0℃ to 50℃
Storage Temperature−20℃ to 60℃
Humidity5% to 95% RH, no condensation
Sensor Cable Length2 meters
Inlet/Outlet Interface6mm inlet, 10mm outlet and drain
Measurement UnitNTU
Range0-1NTU, 0-20NTU, 0-100NTU (configurable)
Accuracy±2% or ±0.015NTU (larger value), based on Formazin standard at 25°C
Zero Offset≤±0.015NTU
Resolution0.001NTU
Water Flow Rate50ml/min to 300ml/min
Sensor Dimensions66mm*145mm*54mm (W*H*D)

Frequently asked questions

What is the measurement range of the turbidity analyzer?

The measurement range is 0-1NTU, 0-20NTU, or 0-100NTU, depending on the model configuration.

What is the accuracy of the turbidity measurement?

The accuracy is ±2% or ±0.015NTU, whichever is larger, based on Formazin standard at 25°C.

What communication interfaces does the analyzer support?

The analyzer supports both Modbus RTU (RS485) and 4-20mA interfaces for data output.

What is the power supply requirement?

The analyzer requires a DC24V power supply.