This turbidity online analyzer is a patented product developed specifically for drinking water quality online monitoring. It has the characteristics of ultra-low turbidity detection limit, high precision measurement, long-term maintenance-free equipment, water-saving work and digital output, supporting cloud platform and mobile terminal data remote monitoring, as well as RS485-modbus communication. It can be widely used in tap water factory water, secondary water supply, pipeline end water, drinking water, membrane filtered water, swimming pool, surface water and other turbidity online monitoring.
This user manual provides detailed instructions for operating and maintaining the Sinomeasure U-YZDPTU300-LCCN1 turbidity online analyzer. It covers safety information, technical specifications, mechanical installation, operational procedures, calibration methods, and data interfaces including Modbus RTU and 4-20mA. The device uses laser scattering to measure water turbidity and supports remote monitoring via cloud and mobile platforms.
The analyzer is designed for continuous monitoring of water turbidity in applications such as drinking water, swimming pools, and surface water. It features IP54 protection, a 24V DC power supply, and supports multiple measurement ranges (0-1NTU, 0-20NTU, 0-100NTU). Maintenance tasks include checking water flow, sensor condition, and cleaning the chamber. This manual is written in Chinese.
| Detection Method | Laser 90° scattering |
|---|---|
| Main Standard | JJG880-2006 |
| Material | Composite material |
| Dimensions | 183mm*349mm*113.5mm (W*H*D) |
| Weight | 4.5Kg |
| Power Supply | DC24V |
| Protection Rating | IP54 |
| Installation Type | Wall-mounted |
| Operating Temperature | 0℃ to 50℃ |
| Storage Temperature | −20℃ to 60℃ |
| Humidity | 5% to 95% RH, no condensation |
| Sensor Cable Length | 2 meters |
| Inlet/Outlet Interface | 6mm inlet, 10mm outlet and drain |
| Measurement Unit | NTU |
| Range | 0-1NTU, 0-20NTU, 0-100NTU (configurable) |
| Accuracy | ±2% or ±0.015NTU (larger value), based on Formazin standard at 25°C |
| Zero Offset | ≤±0.015NTU |
| Resolution | 0.001NTU |
| Water Flow Rate | 50ml/min to 300ml/min |
| Sensor Dimensions | 66mm*145mm*54mm (W*H*D) |
The measurement range is 0-1NTU, 0-20NTU, or 0-100NTU, depending on the model configuration.
The accuracy is ±2% or ±0.015NTU, whichever is larger, based on Formazin standard at 25°C.
The analyzer supports both Modbus RTU (RS485) and 4-20mA interfaces for data output.
The analyzer requires a DC24V power supply.