Yokogawa TDLS220 Tunable Diode Laser Spectroscopy Analyzer User Manual

Describes the use, precautions, maintenance and maintenance of the TDLS220 tunable diode laser analyzer


Yokogawa TDLS220 Tunable Diode Laser Spectroscopy Analyzer User Manual - cover page
Brand
YOKOGAWA
Category
Analyzer
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
1.4 MB
Published
27 October, 2023
Updated
27 October, 2023
MD5 checksum
85161BA26F397054D512E33BFB9C5A24

About this manual

This User's Manual for the Yokogawa TDLS220 Tunable Diode Laser Spectroscopy Analyzer is the Start-up Manual IM11Y01B02-11E-A, 3rd Edition from August 14, 2012, published by Yokogawa Corporation of America. It is for owners and operators of the Model TDLS220 and explains the functions in the product, along with instrument check, quick start, installation, and wiring guidance.

The manual includes safety precautions for the human body, product, and system, covering danger, warning, caution, and note symbols. It also describes safe lifting and carrying, electrical hazard, thermal hazard, chemical hazard, labels, warranty, and service information.

What's inside

  • Safety Precautions
  • Instrument Check
  • Quick Start
  • Installation and Wiring
  • Mounting the Analyzer
  • Sample Inlet and Outlet Considerations
  • Wiring Details
  • Purge Gas Requirements and Hazardous Area Systems
  • Revision Record

Specifications

Power~220v, 50 hz or ~120v , 60 hz (model specific)
Gas cell maximum temperature+120 ºC

Frequently asked questions

How should the Yokogawa TDLS220 be lifted and carried?

It should be lifted by at least two people wearing protective gloves and steel toe boots. It should be transported vertically or horizontally with the mounting plate facing down.

What should be done before disconnecting the analyzer gas tubes?

The analyzer should be flushed with nitrogen or instrument air for 15 minutes before disconnecting from the gas lines. Check for leaks after installation before supplying the gas sample.

What electrical precautions apply when using the TDLS220?

The analyzer is powered by either a ~220v, 50 hz or ~120v , 60 hz (model specific). Disconnect wires from any voltage sources during procedures, and connect protective ground wire to designated terminal.