Describes the use, precautions, maintenance and maintenance of the TDLS220 tunable diode laser analyzer
This User's Manual for the Yokogawa TDLS220 Tunable Diode Laser Spectroscopy Analyzer is the Start-up Manual IM11Y01B02-11E-A, 3rd Edition from August 14, 2012, published by Yokogawa Corporation of America. It is for owners and operators of the Model TDLS220 and explains the functions in the product, along with instrument check, quick start, installation, and wiring guidance.
The manual includes safety precautions for the human body, product, and system, covering danger, warning, caution, and note symbols. It also describes safe lifting and carrying, electrical hazard, thermal hazard, chemical hazard, labels, warranty, and service information.
| Power | ~220v, 50 hz or ~120v , 60 hz (model specific) |
|---|---|
| Gas cell maximum temperature | +120 ºC |
It should be lifted by at least two people wearing protective gloves and steel toe boots. It should be transported vertically or horizontally with the mounting plate facing down.
The analyzer should be flushed with nitrogen or instrument air for 15 minutes before disconnecting from the gas lines. Check for leaks after installation before supplying the gas sample.
The analyzer is powered by either a ~220v, 50 hz or ~120v , 60 hz (model specific). Disconnect wires from any voltage sources during procedures, and connect protective ground wire to designated terminal.