This manual introduces the specifications, safety information, overview, installation guidelines and other contents of Sigma SD900 refrigerated sampler and all weather refrigerated sampler.
This manual provides detailed instructions for installing, operating, and maintaining the HACH SD900 Refrigerated Sampler and All Weather Refrigerated Sampler. It includes sections on safety, component overview, electrical and tubing installation, controller operation, sampling programs, volume calibration, and advanced sampling features. Maintenance procedures cover pump, distributor arm, and desiccant replacement, as well as troubleshooting and diagnostic tests. Specifications for both sampler types are provided, including dimensions, power requirements, and environmental operating conditions.
| Refrigerated Sampler Dimensions | Height: 112 cm (44 in.), Width: 61 cm (24 in.), Depth: 61 cm (24 in.), Weight: 63.3 kg (140 lb) |
|---|---|
| Refrigerated Sampler Power Requirements | 115 VAC, 60 Hz, 3.3 A (18 locked rotor amps) or 230 VAC, 50 Hz, 1.7 A (9 locked rotor amps) |
| All Weather Refrigerated Sampler Dimensions | Height (lid closed): 130 cm (51 in.), Height (lid open): 180 cm (71 in.), Width: 76 cm (30 in.), Depth: 81 cm (32 in.), Weight: 86 kg (190 lb) |
| All Weather Refrigerated Sampler Power Requirements | 115VAC, 60Hz, 4.2A or 6.4A with optional controller compartment heater or 230VAC, 50Hz, 2.7A or 4.1A with optional controller compartment heater |
| Sample Bottle Capacity (Single Bottle Mode) | 21 L (5.5 gal) polyethylene or 10 L (2.5 gal) polyethylene or 10 L (2.5 gal) glass bottle |
| Pump Flow Rate | 1.25 gpm (4.8 L/min) at 3 ft (1 m) vertical lift using 3/8-in. intake tube |
| Pump Tubing Life | 20,000 sample cycles under specified conditions |
| Maximum Vertical Lift | Minimum of 28 ft, using 29 ft of 3/8-in. vinyl intake tube at sea level at 20–25 °C (68–77 °F) |
| Operating Temperature (Refrigerated Sampler) | 0–50 °C (32–122 °F) |
115 VAC, 60 Hz, 3.3 A (18 locked rotor amps) or 230 VAC, 50 Hz, 1.7 A (9 locked rotor amps).
Minimum of 28 ft, using 29 ft of 3/8-in. vinyl intake tube at sea level at 20–25 °C (68–77 °F).
1.25 gpm (4.8 L/min) at 3 ft (1 m) vertical lift using 3/8-in. intake tube.
From -40 to 50 °C (-40 to 122 °F).