M-Series® Mag Meter Model M-5000 Installation & Operation Manual
The Badger Meter M-5000 is an electromagnetic flow meter designed for fluid metering in various industries, including water, wastewater, food and beverage, pharmaceutical, and chemical applications. The system consists of a detector, which includes the flow tube, isolating liner, and measuring electrodes, and an amplifier responsible for signal processing and display. The manual details the setup of both integral meter mount and remote junction box configurations.
| Flow Range | 0.328 to 32.8 fps (0.1 to 10 m/s) |
|---|---|
| Sizes | 1/2-inch to 24 inches (15 mm to 600 mm) |
| Min. Conductivity | ≥ 20 micro siemens/cm |
| Accuracy | ± 0.50% of rate for velocities greater than 1.64 ft/s |
| Electrode Materials | Standard: Alloy C; Optional: 316 stainless steel |
| Liner Material | PTFE 1/2 inch to 24 inches, hard rubber 1 inch to 24 inches |
| Fluid Temperature (Remote Amplifier) | PTFE 311°F (155°C); Hard Rubber 178°F (80°C) |
| Ambient Temperature | -4°F to 140°F (-20°C to 60°C) |
| Meter Housing Material | Carbon steel welded |
| Flanges | Carbon steel - Standard (ANSI B16.5 Class 150 RF); 316 stainless steel (optional) |
An equivalent of three diameters of straight pipe is required on the inlet side, and two diameters are required on the outlet side.
Yes, mag meters can operate accurately in any pipeline orientation and can measure volumetric flow in forward and reverse directions.
The minimum conductivity required is ≥ 20 micro siemens/cm.
The standard electrode material is Alloy C, with 316 stainless steel available as an optional material.