Introduced the importance of dynamic pressure calibration standard, including existing standards, industry users and related companies, introduced Kistler's pressure calibration instruments, and gave the specific parameters of the products
This presentation was given at the 8th International Workshop Analysis of Dynamic Measurements by Mark Schiefer. It argues that a large user community exists for dynamic pressure calibration and that no ties to national metrology institutes currently exist, then lists the manufacturers of commercial excitation devices and the existing ISA-37-16-01-2002A guide for dynamic calibration of pressure transducers, along with the PIWG users group led by companies such as Kulite, Meggitt, Siemens, GE, Boeing, PW, Rolls-Royce, Alstom, MTU, Honeywell, Dytran and Kistler.
The slides then describe a range of calibration workstations with their maximum pressures, input types and rise times, followed by a proposed outline for a dynamic pressure standard covering scope, normative references, terms and definitions, uncertainty of measurement, apparatus (shock tube, low and high pressure pneumatic and hydraulic systems), ambient conditions, preferred rise time durations, methods for different sensor types and reporting of calibration results. It closes with the standardization efforts in ISA 107.5 and ISO TC108 SC3 WG6 and the author's contact details.
| K9903C maximum pressure | 10 bar |
|---|---|
| K9903C rise time | 3 to 5 ms |
| K9907C maximum pressure | 69 bar |
| K9913C maximum pressure | 1034 bar |
| K9905C maximum pressure | 6905 bar |
| K9905C input | Quasi-static input |
| Kistler 6906 calibrator range | 10,000 bar |
| Kistler 6909 calibrator range | 5,000 bar |
The document cites ISA-37-16-01-2002A, A Guide for Dynamic Calibration of Pressure Transducers.
Coverage would be by excitation device: shock tube, low pressure gas, high pressure gas, low pressure hydraulic and high pressure hydraulic.
The K9913C has a maximum pressure of 1034 bar, with impulse (hammer) input and silicon oil media.
ISO TC108 SC3 WG6 created a preliminary work item in 10/2013, with Schiefer chosen as group leader.