This manual introduces the function and use of the MultiCal Hg/PSI Pressure Module.
The MultiCal Hg/PSI Pressure Module measures pneumatic or hydraulic pressure, and vacuum, by converting pressure into an electrical output. It has no display: the module outputs 1 millivolt per pressure unit (inches Hg or psi) and is read with a device that measures DC millivolts, such as a digital multimeter or an oscilloscope. A three-position switch turns the unit on and selects inches Hg or psi scaling, and the OFF position is used to check the internal battery condition through an external meter. Pressure is applied through one of two 1/8" female NPT ports while any unused port is plugged.
This operation manual covers the module's functions, the pressure measurement procedure, battery replacement, recommended calibration equipment and the calibration procedure, and the pressure, operating environment and general specifications. It includes warnings not to exceed the pressure limits of tubing and fittings and not to insert objects into the inlet ports, plus an EC declaration of conformity, factory service information, trademarks and warranty terms. Crystal Engineering is named as the company that designs, manufactures, markets and services the MultiCal; the cover carries the AMETEK Test & Calibration Instruments mark.
| Output scaling | 1 millivolt per pressure unit (inches Hg or psi); vacuum readings are negative |
|---|---|
| Accuracy, 0 to 30 inHg | ±(0.1% of Reading + 0.01 inHg) |
| Accuracy, 0 to 500 psi | ±(0.1% of Reading + 0.02 psi) |
| Accuracy, 0 to -14.7 psi (vacuum) | ±(0.25% of Reading), typical |
| Wetted materials | 316 stainless steel, PTFE impregnated hard anodized aluminum, and buna-n rubber |
| Pressure ports | Two 1/8" female NPT ports |
| Meter input impedance | 10 MΩ or greater |
| Humidity, 10 to 30°C | 0 to 95% relative humidity |
| Recommended calibration cycle | 1 year |
Any meter with 10 MΩ or greater input impedance, such as most handheld multimeters and some benchtop multimeters, set to the DC millivolt range.
With the switch still in the OFF position, the multimeter must indicate a minimum of 100 mV; readings below that mean the battery must be replaced.
A calibration cycle of 1 year is recommended to keep the module within specifications.
Two ports are provided so extra tees are not needed; plug any unused ports.