This is a Photosynthetic Active Radiometer PAR lite user manual
The Kipp & Zonen PAR LITE is an instrument for measuring Photosynthetic Photon Flux Density, also called PAR, which represents the number of photons between 400 and 700 nm incident per square meter per second. It is designed for continuous outdoor use and is typically used in agrometeorological and horticulture applications.
The manual covers general information, a five-minute user guide, sensor properties, calibration, installation and maintenance, troubleshooting, delivery contents, and accessories. It also provides electrical and spectral specifications, including nominal impedance, sensitivity, and spectral range.
| Nominal impedance | 240 Ω |
|---|---|
| Response time | < 0.1 s |
| Sensitivity | 4 - 6 µV/ µmol/s.m2 |
| Expected signal range | 0 – 12 mV |
| Non-stability | within ± 2%/year |
| Non linearity | within ± 1% up to 10000 µmol /s.m2 |
| Temperature dependence of sensitivity | within ± 0.2%/°C (-0.1% typ.) |
| Detector type | Silicon photodiode plus filters |
| Spectral range | 400 - 700 nm |
| Cut on (50 %) | 400 ± 15 nm |
| Cut off (50 %) | 700 ± 15 nm |
It measures Photosynthetic Photon Flux Density, also called PAR, which is the number of photons between 400 and 700 nm incident per square meter per second.
Recalibration is suggested every two years, preferably by comparing daily totals on a clear day and a cloudy day.
Delivery includes one PAR LITE sensor, one calibration certificate, and one manual.