Vapotherm Hi-VNI Mask-Free NIV Pediatric Pocket Guide

This file is the Vapotherm Pediatric Pocket Guide guidelines and best practices for mask-free NIV™ technology for spontaneously breathing patients


Vapotherm Hi-VNI Mask-Free NIV Pediatric Pocket Guide - cover page
Brand
Vapotherm
Category
Professional medical equipment
Document type
Other
Language
English
Pages
8
File format
PDF
File size
1.1 MB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
BC99C03C16D3FBF58A55E06A58FC37C7

About this manual

The Vapotherm Pediatric Pocket Guide presents guidelines and best practices for mask-free NIV with Vapotherm Hi-VNI Technology in spontaneously breathing pediatric patients. It lists patient selection signs and diagnoses, Hi-VNI cannula sizes and flow ranges, and instructions for fitting and applying the cannula with the Precision Flow system.

The guide also covers setting temperature and FiO2, starting and titrating flow, monitoring patient and device parameters, and weaning flow or FiO2. Separate sections address use with aerosol medication, nitric oxide delivery systems, and Precision Flow Heliox, including compatible disposable patient circuits and flow ranges.

What's inside

  • Patient Selection
  • Diagnoses
  • Hi-VNI Cannula Selection
  • Fitting the Hi-VNI Cannula
  • Hi-VNI Cannula Application
  • Therapy Implementation and Maintenance
  • Monitoring Therapy
  • Weaning
  • Aerosol Medication and Specialty Gases
  • Use with Precision Flow Heliox

Specifications

Pediatric Small cannula flow range1-20 L/min
Pediatric/Adult Small cannula flow range and tip OD5-40 L/min; 2.7 mm tip OD
Adult cannula flow range and tip OD5-40 L/min; 4.8 mm tip OD
PF-DPC-LOW flow range1-8 L/min
PF-DPC-HIGH flow range5-40 L/min
Recommended starting flow for patients under 20 kg2 L/min per kilogram of weight
Set temperature37°C

Frequently asked questions

What starting flow rate is recommended for infants and toddlers under 20 kg?

The recommended starting rate is 2 L/min per kilogram of weight, then titrate to clinical effect.

Which cannulae should be used with the Vapotherm Precision Flow?

Only Vapotherm Hi-VNI cannulae should be used with the Vapotherm Precision Flow.

What can be done if rainout occurs at less than 10 L/min?

Consider dropping the temperature to no lower than 34°C.

What is the Precision Flow operational flow range with PF-DPC-HIGH?

With PF-DPC-HIGH (Blue packaging), the flow range is 5-40 L/min.

Download PDF (1.1 MB)