Skeletal Dynamics ALIGN Radial Head System Surgical Technique Guide

ALIGN® Radial Head System Surgical Technique Guide


Skeletal Dynamics ALIGN Radial Head System Surgical Technique Guide - cover page
Brand
skeletal dynamics
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
24
File format
PDF
File size
2.2 MB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
0A26891EA5B8F452B9BAECCA6F7EE0B9

About this manual

The guide covers indications for use, including primary replacement after fracture, revision, and symptomatic sequelae after resection. It details surgical steps from landmark identification and exposure to head and neck sizing, resection, canal preparation, and planing.

The document also provides a complete instrument tray listing and implant catalog with specifications for cobalt-chromium heads (18–26 mm) and titanium stems (7–11 mm diameter, 0–8 mm offset), all designed for press-fit fixation with titanium plasma spray coating.

What's inside

  • Indications for Use
  • Elbow Landmarks
  • Distal Ulna Landmarks
  • Superficial Exposure
  • Deep Exposure
  • Head Sizing
  • Trial Head Selection
  • Provisional Neck Sizing
  • Initial Neck Sizing
  • Resection

Specifications

Head diameters18mm-26mm
Neck offsets0mm-8mm
Stem diameters7mm-11mm
Head materialCoCr
Stem materialTi
Surface coatingTi plasma spray
Fixation methodpress-fit
Maximum defect correction23mm

Frequently asked questions

What are the indications for the ALIGN Radial Head System?

The system is indicated for replacement of the radial head for degenerative or post-traumatic disabilities with pain, crepitation, decreased motion, joint destruction, subluxation, resistance to conservative treatment, primary replacement after fracture, symptomatic sequelae after resection, and revision following failed arthroplasty.

What head diameters are available for the ALIGN Radial Head System?

The system offers five head diameters ranging from 18 mm to 26 mm.

What is the maximum radial neck defect that can be corrected with this system?

The maximum defect that can be corrected is 23 mm.