The HI99151 is a waterproof, portable pH and temperature meter designed for use in beer production. It features automatic calibration, a dual-line display, and a rugged FC2143 probe with integrated temperature sensing. The device supports one or two-point calibration using standard or NIST buffers and includes battery life indication and auto-off functionality. It is IP67 rated and suitable for field use.

The manual covers setup, operation, electrode maintenance, calibration, troubleshooting, and warranty information. It includes detailed instructions for connecting the probe, performing measurements, freezing readings, and maintaining the electrode for optimal performance in brewing environments.

What's inside

  • Preliminary Examination
  • General Description and Intended Use
  • Main Features
  • Specifications
  • Display Description
  • Operational Guide
  • Meter Setup
  • Battery Replacement
  • Accessories
  • Electrode Maintenance

Specifications

Range‑2.00 to 16.00 pH / ‑2.0 to 16.0 pH ±825 mV (pH‑mV) ‑5.0 to 105.0 °C/23.0 to 221.0 °F
Resolution0.01 pH / 0.1 pH 1 mV 0.1 °C/0.1 °F
Accuracy @ 25 °C / 77 °F±0.02 pH / ±0.1 pH ±1 mV (pH‑mV) ±0.5 °C up to 60 °C; ±1.0 °C outside ±1.0 °F up to 140 °F; ±2.0 °F outside
Temperature CompensationAutomatic ‑5.0 to 105.0 °C/23.0 to 221.0 °F
pH CalibrationAutomatic, 1 or 2 point selectable buffer set Standard (4.01, 7.01, 10.01) or NIST (4.01, 6.86, 9.18)
Probe (included)FC2143 preamplified pH/temperature probe with flat glass tip, DIN connector and 1 m (3.3’) cable
Battery type/life1.5V AAA (3 pcs.) approx. 1400 hours of continuous use
Environment0 to 50 °C (32 to 122 °F) RH max. 100%
Meter dimensions154 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Weight (with batteries)196 g (6.91 oz.)
Case ingress protection ratingIP67

Frequently asked questions

What is the pH measurement range of the HI99151?

The pH measurement range is ‑2.00 to 16.00 pH.

What type of probe is included with the HI99151?

The HI99151 includes an FC2143 preamplified pH/temperature probe with a flat glass tip, DIN connector, and 1 m (3.3’) cable.

How long does the battery last?

The battery life is approximately 1400 hours of continuous use with 3 x 1.5V AAA batteries.

What is the warranty for the HI99151?

The HI99151 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. Electrodes and probes are warranted for six months.