This document provides information on how to set up printers with the QIAstat-Dx Analyzer 1.0. It includes information on recommended printers, general printer use recommendations, USB printer setup, network printer setup, Windows 7 PC direct connection setup, and Windows 10 PC direct connection setup.
This guide provides instructions for configuring printers with the QIAstat-Dx Analyzer 1.0, including recommended printer models and setup steps for USB, network, and direct PC connections. It also covers troubleshooting common printer issues and CUPS configuration.
The guide includes detailed steps for setting static IP addresses on both the analyzer and Windows PCs, as well as guidance on using generic printer drivers and troubleshooting print jobs that fail or do not appear in the application software.
The CUPS password is case sensitive. If a particular CUPS password does not work despite being entered correctly, disable CUPS, save the option, re-enable CUPS and save the option to get a new password.
You can find the IP address of your QIAstat-Dx Analyzer under Settings Network IPv4 Settings IPv4 Address.
Try to install a more specific driver for your printer via CUPS. For this purpose, follow the instructions for setting up network printers (see above). Ensure the QIAstat-Dx Analyzer and connecting PC are part of the same local network (see “Network settings” of the QIAstat-Dx Analyzer 1.0 User Manual).
Ensure that the printer is powered on and that the USB or network connection is established successfully. Your printer can be accessed via a generic printer driver that is pre-installed on the QIAstat-Dx Analyzer. Try to print via this driver (see “Printer settings” in the QIAstat-Dx Analyzer 1.0 User Manual). If a custom printer driver was already previously configured, power OFF the QIAstat-Dx Operational Module and then power ON again to make the driver available. To power OFF or ON, use the switch at the back of the instrument. The QIAstat-Dx application software detects the available printer drivers upon start up. When adding a new printer, the QIAstat-Dx Operational Module must be restarted before the new driver will be available.