University of Cincinnati Rodent Anesthesia and System Inspection Guideline

This guideline provides general recommendations for anesthesia and inspection of anesthesia systems of laboratory rodents at the University of Cincinnati.


University of Cincinnati Rodent Anesthesia and System Inspection Guideline - cover page
Brand
University of Cincinnati
Category
Other
Document type
Other
Language
English
Pages
4
File format
PDF
File size
168 KB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
1057A9951F008CA70CDE2A50033113F8

About this manual

This document outlines general recommendations for the anesthesia and inspection of anesthesia systems for laboratory rodents at the University of Cincinnati. It details considerations for performing anesthesia, including acclimation periods, animal characteristics, and specific handling for corneal drying and hypothermia. The text emphasizes the use of inhalation anesthesia, particularly isoflurane with a non-rebreathing system, and describes the drop box method for ultra-short procedures or euthanasia.

What's inside

  • Considerations When Performing Anesthesia
  • Anesthetic Delivery Mechanisms
  • Anesthesia monitoring and recovery
  • Waste Anesthetic Gas Canister Use, Placement and Disposal
  • Anesthesia and Vaporizer Inspection Guidelines

Frequently asked questions

What is the preferred anesthetic for both short and long procedures in rodents?

Isoflurane is preferred for both short and long procedures.

What is the maximum weight increase allowed for a 50 gm waste anesthetic gas canister before it must be replaced?

The canister must not exceed 50 gm more than the initial baseline weight.

How often must anesthesia systems be inspected by a qualified professional?

Anesthesia systems must be inspected annually.

What is the required action if a vaporizer is found to be ± 15% out of calibration?

The vaporizer must be serviced off-site by qualified personnel or company.