WatchGas PDM Single Gas Detector User Manual

The PDM is a portable multigas monitor designed to warn the user about the presence of dangerous concentrations of oxygen, combustible, or toxic gases in the atmosphere.


WatchGas PDM Single Gas Detector User Manual - cover page
Brand
WatchGas
Document type
User Manual
Language
English
Pages
14
File format
PDF
File size
932 KB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
26E825719AFB426BDE684D0BB57CA978

About this manual

The WatchGas PDM Single Gas Detector is a portable multigas monitor designed to alert users to dangerous gas concentrations. It features real-time gas detection, alarms via sound, vibration, and flashing LEDs, and a display showing concentration levels. The device is activated by holding the key for three seconds and requires regular calibration and maintenance. It is not repairable or modifiable, and its battery and sensor are disposable. Calibration and self-test procedures are detailed, including fresh air and standard gas calibration, and bump tests are configurable via IR-Link.

Safety warnings emphasize proper use, regular testing, and avoiding unauthorized modifications. The device displays alarm thresholds for gases like oxygen, CO, H2S, and others, with default values based on international standards. Users must ensure clean air for calibration and avoid using the device beyond its 24-month lifespan.

What's inside

  • Product Overview
  • Activation
  • Mode
  • Alarms / Battery / Test Failure Display
  • Calibration
  • Self Test & Bump Test
  • Specifications
  • Certificates
  • Limited Warranty

Specifications

Gas Types DetectedOxygen, Combustible, Toxic Gases
Alarm IndicatorsAcoustic alarm, flashing LEDs, vibration
Display UnitsOxygen: %VOL, Toxic Gases: ppm
Calibration FrequencyFresh air calibration every 30 days

Frequently asked questions

How do I activate the PDM?

Press and hold the key for three seconds. After a 10-second countdown, the device will stabilize and enter detection mode.

What happens if calibration fails?

The device will display a failure message and beep intermittently. Check for contaminants or empty gas cylinders and retry. If it fails repeatedly, contact WatchGas.

Can I change the alarm set points?

Only authorized experts may change the alarm set points, as they are set according to international standards.

How often should I perform a fresh air calibration?

At least every 30 days, in clean air without contaminants.