This manual describes the use of the iBreeze CPAP system, including system composition, safety information, operation, cleaning and maintenance.
This manual provides instructions for the Resvent iBreeze CPAP Systems, which include the iBreeze 20C, iBreeze 20C Pro, iBreeze 20A, and iBreeze 20A Pro models. These devices deliver continuous positive airway pressure (CPAP) and auto-adjusting positive airway pressure (APAP) therapy for the treatment of Obstructive Sleep Apnea (OSA) in spontaneously breathing patients weighing over 30kg (66lbs). The guide is intended for use in home, hospital, or institutional environments and must be used in accordance with the instructions provided by a health care professional.
The document covers essential topics including intended use, contraindications, safety information, system setup, operation, device alerts, troubleshooting, and cleaning procedures. It emphasizes that the device is not a life-supporting device and should only be used with masks and accessories recommended by Resvent to ensure safety and proper performance.
| Indication | treatment of Obstructive Sleep Apnea (OSA) |
|---|---|
| Patient Weight Requirement | over 30kg (66lbs) |
| Maximum Therapeutic Pressure | 20 cmH2O |
| Maximum Fault Pressure | 40 cmH2O |
| Maximum Water Fill Level | 290 mL |
| Warm-up Time | at least 1h |
| Cool-down Time | at least 1h |
| Maximum Ambient Temperature | 35°C (95°F) |
The iBreeze CPAP Systems include the iBreeze 20C, iBreeze 20C Pro, iBreeze 20A, and iBreeze 20A Pro.
Resvent is responsible for the effects on safety, reliability and performance of this product, provided installation and repairs are conducted by authorized personnel.
Yes, but the device must be turned on and airflow generated before the oxygen supply is turned on, and the oxygen supply must be turned off before the device is turned off.
The maximum fill level is 290 mL.