This report is an instruction manual in the set-up, calibration, and operation of the Panametrics Epoch 2002 ultrasonic testing workstation. Detailed procedures for sample preparation and measurement of different types of ice are presented.
This guide provides step-by-step instructions for operating the Panametrics Epoch 2002 ultrasonic testing workstation, specifically for measuring ultrasonic velocity in glacier ice. It covers setup, calibration procedures, sample preparation, and measurement techniques. The Epoch 2002 is a portable system with a digital oscilloscope, pulser/receiver controls, and microprocessor-based functions. It can operate in manual, external, or external-program modes via RS-232. The manual assumes manual mode and includes details on keypad layout, display settings, and troubleshooting. It also notes that the unit has been modified for low-temperature operation and may drift above 32–38°C.
The Epoch 2002 is designed for generating, detecting, and processing ultrasonic signals using transducers. In ice measurements, it calculates velocity from observed travel time using a known path length. The manual references manufacturer specifications (Table 1) and includes figures for keypad layout and display screens. It also mentions that the unit can propagate and detect a 2.25MHz P-wave signal through up to 20–25 cm of polycrystalline ice.
| Display Graticule | 2.25" x 3.125" (57 x 80mm) - approximate size |
|---|---|
| Velocity Range | 0.025 to 0.6 in / jis (635 to 15240m/s) |
| Operating Temperature | -20°C (+25°C typical) to 50°C |
| Weight | 12.5 Ibs. (5.7 Kg) without battery |
| Battery Life | 12 hrs. @ 25°C |
| Transducer Connector | BNC or Number 1 Lemo |
The operating temperature range is -20°C (+25°C typical) to 50°C.
The battery life is 12 hours at 25°C.
The Epoch 2002 supports manual mode (via soft keypad), external mode (via RS-232 from remote computer), and external-program mode (via RS-232 with a program).