This document is the AutoVent 4000CPAP Pneumatically Powered Ventilator & CPAP Generator operation manual, which introduces the characteristics and usage of this product.
This manual provides step-by-step instructions for operating the Autovent 4000CPAP ventilator and CPAP generator in both ventilation and CPAP modes. It includes setup procedures, control adjustments, alarm management, and maintenance guidelines. The device is pneumatically powered and designed for emergency respiratory support using oxygen or blended air/oxygen. Models vary by features such as oxygen blending and electronic alarms, with weight and size differences noted for each variant.
| Gas Supply Pressure | 280 kPa (40.6 psi) to 600 kPa (87.0 psi) |
|---|---|
| Breaths per Minute (BPM) Range | 8 to 30 BPM (with 1 sec inspiratory time) |
| Tidal Volume Range | 200 ml to 600 ml (1 sec inspiratory time) |
| Inspiratory Time | 1 second or 2 second selection |
| Safety Pressure Relief | Adjustable from 20 to 80 cm H2O ±10% |
| Weight | L760CPAP = 2.35 kg (5.187 lb) |
| Size | L760CPAP = 73.6 x 203 x 244 mm (2.9 x 8.0 x 9.6 inches) |
| Operating Conditions | -18 to 50°C (0 to 122°F) |
| Oxygen Blending | 65% ±12% oxygen with blender |
| CPAP Range | 4 to 20 cm H2O |
Models include L760CPAP (basic), L761CPAP (with O2 blender), L762CPAP (with O2 blender and electronic alarms), and L763CPAP (basic with electronic alarms).
Models L761 and L762 can blend oxygen to 65% ±12% with a venturi system using ambient air.
Adjustable from 20 to 80 cm H2O, with 60 cm H2O as the maximum regardless of setting.
Operating conditions are -18 to 50°C (0 to 122°F).