This document introduces the IMPACT® alarm system in detail, including its functions, installation and use
This document provides step-by-step installation and operating instructions for the IMPACT® Alarm System, designed to monitor medical gas pressures and vacuums. It covers component descriptions, wiring procedures, module setup, and safety precautions. The system includes modules for pressure, vacuum, and switch signal monitoring, with detailed guidance on installation, wiring, and testing procedures.
The manual includes safety warnings for electrical hazards, gas shutdown requirements, and electrostatic discharge precautions. It also specifies electrical requirements, operating conditions, and accuracy for each module type, ensuring proper installation and safe operation in hospital and laboratory environments.
| Electrical Requirements | AC 115 50/60 Hz, 0.5 Amps Max; AC 230 50/60 Hz, 0.3 Amps Max |
|---|---|
| Operating Temperature | 50°F to 100°F (10°C to 37.8°C) |
| Operating Humidity | 30% to 75% RH |
| Accuracy - 100 PSI Module | ± 3 PSI |
| Accuracy - 300 PSI Module | ± 9 PSI |
| Accuracy - Vacuum Module | ± 1 Inch Hg |
| Accuracy - 3000 PSI Module | ± 60 PSI |
AC 115 50/60 Hz, 0.5 Amps Max; AC 230 50/60 Hz, 0.3 Amps Max.
Operating temperature: 50°F to 100°F (10°C to 37.8°C); Operating humidity: 30% to 75% RH.
Press the ENTER/SILENCE button on the Control Module to silence the audio alarm.
± 3 PSI.