This manual is the instruction manual of the building smoke density tester. It mainly introduces the features, technical parameters and usage methods of the instrument.
This manual provides step-by-step instructions for installing, calibrating, and operating the ZKHW-207 smoke density tester. It covers safety precautions, equipment setup, sample preparation, and testing procedures according to GB/T8627-2007. The device uses PLC + touch screen control with automatic data recording and analysis. Maintenance and troubleshooting are also detailed.
The tester is designed for use in building material testing labs and includes features such as automatic ignition, smoke measurement, and data storage. It supports computer control and network sharing for multiple users. Calibration is required annually or as needed using standard filters. This manual is written in Chinese.
| Model | ZKHW-207 |
|---|---|
| Power Supply | 220V, 50Hz, 100W |
| Control Method | PLC + 5-inch touch screen |
| Smoke Density Range | 0-100% |
| Smoke Density Accuracy | ±3% |
| Burner Type | Bunsen burner, 260mm length, 0.13mm nozzle, 45° tilt |
| Gas Source | Propane or equivalent gas (customer-provided) |
| Test Duration | 4 minutes |
| Data Points | 16 data points at 15s intervals |
| Display Resolution | 800x480 |
| Storage Capacity | 1GB, 1000 test records |
| Standard Filters | 3 standard filters (100%, 75%, 50%, 25%, 0%) |
| Compliance Standard | GB/T8627-2007 |
The smoke density range is 0-100%, with an accuracy of ±3%.
Each test lasts 4 minutes, with data recorded every 15 seconds for a total of 16 data points.
The device uses PLC + 5-inch touch screen control with automatic data recording and analysis.
The device should be calibrated annually, or as needed, using standard filters.