This document is the user manual of 354 series ionization vacuum gauge
This user manual provides detailed instructions for installing, operating, and maintaining the Kurt J. Lesker Company 354 Series Ionization Vacuum Gauge Module. It covers mechanical and electrical installation, setup, analog and RS485 communications, calibration, troubleshooting, and maintenance. The gauge measures vacuum pressure from 1.00 x 10-9 to 5.00 x 10-2 Torr using ionization and features a built-in display, relay, and signal conditioning. Safety precautions are emphasized throughout, including warnings for hazardous voltages and proper grounding.
| Measurement Range | 1.00 x 10-9 to 5.00 x 10-2 Torr / 1.33 x 10-9 to 6.66 x 10-2 mbar / 1.33 x 10-7 to 6.66 Pa |
|---|---|
| Accuracy (N2) | 1.00 x 10-8 to 5.00 x 10-2 Torr; ± 15% of reading |
| Repeatability | ± 5% of reading |
| Display | OLED graphical display, 3 digits plus 2 digits exponent, user-selectable Torr, mbar, or Pa |
| Emission Current | 100 µA, 4 mA, or automatic switching between 100 µA and 4 mA |
| Degas Power | 3 W, electron bombardment |
| Overpressure Protection | gauge turns off at factory default setting of 5.00 x 10-2 Torr |
| Internal Gauge Volume | 1.0 in3 (16.4 cm3) |
| Operating Temperature | 0 to +40 °C |
| Bakeout Temperature | 200 °C (sensor only - electronics removed) |
| Serial Communications | RS485 - ASCII protocol; minimum command interval: 50 ms |
| Analog Output | log-linear 0 to 9 Vdc, 1 V/decade |
| Setpoint Relay | one single-pole, double-throw (SPDT), 1A at 30 Vdc resistive, or ac non-inductive |
| Input Power | 20 to 28 Vdc, 30 W typical |
| Weight | 0.6 lb. (0.27 kg) with NW25 KF flange housing (electronics) |
The measurement range is 1.00 x 10-9 to 5.00 x 10-2 Torr / 1.33 x 10-9 to 6.66 x 10-2 mbar / 1.33 x 10-7 to 6.66 Pa.
The accuracy for nitrogen is ±15% of reading within the range 1.00 x 10-8 to 5.00 x 10-2 Torr.
The gauge turns off at the factory default setting of 5.00 x 10-2 Torr.
The operating temperature range is 0 to +40 °C.