KJL-6000 series thermocouple vacuum gauge tube operator's manual, including the product specifications, accuracy, response time, baking temperature, positive pressure rating, mounting orientation, standard tube materials, troubleshooting procedure, troubleshooting notes, etc.
This document serves as the operator's manual for the KJL-6000 series thermocouple vacuum gauge tube manufactured by the Kurt J. Lesker Company. It details the device's specifications, including a pressure range from 1 x 10-3 to 1 Torr, and describes the operating principles based on heat transfer from a heated filament to gas molecules. The manual also outlines the standard materials used, such as stainless steel for the cup and stem, and nickel-plated Kovar for connector pins.
| Pressure Range | 1 x 10-3 to 1 Torr |
|---|---|
| Accuracy | ± 1 mTorr from 1 to 20 mTorr |
| Response Time | < 1 second for 1 mTorr up to atmosphere |
| Bakeout Temperature | 100 °C (212 °F) |
| Positive Pressure Rating | Tested to 1200 PSIG for operation up to 300 PSIG |
| Standard Tube Materials | Stainless steel (cup, stem and flange) |
Use the ohmmeter setting on a multimeter to measure the resistance between pins 3 and 5. A reading of 19 ohms +/- 2 ohms indicates the tube is good.
Ideally, the open end should point down so as to be self-draining should any vapors condense.
It is a good idea to allow the tube to outgas in the vacuum system for approximately 24 hours, but allow at least 15 minutes for the tube to stabilize.
No, thermocouple gauges should definitely not be used to monitor the backfilling of loadlocks as it is possible to create a dangerous overpressure situation.