This manual provides guidance on the secondary operations of medical device production, including assembly, packaging, labeling, and other operations.
This guide details methods for joining Eastman Tritan ™ copolyester, including solvent bonding, adhesive bonding, ultrasonic welding, and heat staking. It covers specific joint designs like energy directors and step joints for optimal bond strength. Thermal methods such as RF and laser welding are also described, with warnings and recommendations for safe use. The guide emphasizes consulting Eastman technical service for medical device design and bonding applications.
Methyl ethyl ketone (MEK), cyclohexanone, tetrahydrofuran, 50/50 cyclohexanone/methylene chloride, and 50/50 to 80/20 MEK/cyclohexanone.
A recommended starting temperature is ~260°C (~500°F), adjusted until the material starts to stick to the horn, then reduced by 6°–8°C (10°–15°F).
The included angle of the energy director should be 60°.