Negele Messtechnik MPI-300 NCS-M Parameter Tool User Manual

This document is the MPI-300 NCS-M Parameter Tool user manual, which introduces the use of the product, including system requirements, applicable range, product size, electrical connection, technical data, installation of USB driver, preparation of parameter tool, user interface, start of application, connection with sensor, display of sensor data, sensor status, configuration of sensor, reading and storing configuration, disconnect sensor, etc.


Negele Messtechnik MPI-300 NCS-M Parameter Tool User Manual - cover page
Brand
Negele Messtechnik
Category
Health Care
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
314 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
1E396DBB0ED4E33BB08FB4A559CA1A61

About this manual

This document provides instructions for using the MPI-300 NCS-M Parameter Tool to configure NCS-M capacitive level sensors. It covers system requirements, installation of USB drivers, connecting the sensor, and adjusting parameters such as switching threshold, hysteresis, and time delay. The tool is designed for use with Windows operating systems and requires a USB 2.0 port. Safety precautions and technical specifications for the NCS-M sensor are also included.

The manual explains how to start the application, connect to the sensor, display sensor data, configure settings, read and store configurations, and disconnect the sensor. It includes technical data for the NCS-M sensor, such as materials, temperature ranges, and electrical connections. The document also outlines conformity with EU standards and safety guidelines for installation and maintenance.

What's inside

  • Application
  • System Requirements
  • Conformity with Standards
  • NCS-M Capacitive Level Sensor
  • MPI-300 Programming Adapter
  • NCS-M Parameter Tool
  • User Interface
  • Operation of the Parameter Tool
  • Connecting with the Sensor
  • Display of Sensor Data

Specifications

Process connectionThread G½" CLEANadapt combined with Anderson-Negele weld-in sleeves, build-in systems and adapter sleeves
Tightening torqueMax. 10 Nm
Surface qualityR a < 0,8 µm
MaterialsThreaded connector: Stainless steel 1.4305, Coupling part: PEEK (FDA approval: 21 CFR 177 2415)
Temperature rangesAmbient: 0…70 °C, Process: -10…115 °C, CIP / SIP: 143 °C max. 120 min
Electrical connectionM12 connector, stainless steel 1.4305, 16…32V DC, IP 69K
OutputPNP active 50 mA (short-circuit proof)
Measurement rangeNCS-M-11: Dk > 20, NCS-M-12: Dk > 2
Weight100 g

Frequently asked questions

What operating systems are supported by the MPI-300 NCS-M Parameter Tool?

The tool supports Windows XP (SP3), Vista (SP1), Win 7, Win 8, and Win 10.

What is the minimum CPU requirement for the MPI-300 NCS-M Parameter Tool?

The minimum CPU requirement is 1 GHz.

What is the switching threshold for the NCS-M-12 sensor?

The switching threshold for the NCS-M-12 sensor is 1%.

How do I install the USB driver for the MPI-300 Programming Adapter?

Insert the CD, double-click the setup file in the "MPI-300 USB" directory, follow the installation assistant, and the USB adapter will be detected.