This document is the instruction manual for the HC1 series electro-conductivity sensors produced by Anderson Instrument Co. Inc.
This manual provides installation and operational guidance for Anderson Instrument Co. Inc.'s HC1 Series electrodeless conductivity sensors. It details mounting options, wetted materials (PVDF or PFA Teflon), operating precautions, and termination procedures for sensor and interconnect cables. The sensor measures conductivity from 0-200 to 0-2,000,000 microSiemens/cm and is constructed with 3A-approved materials for sanitary applications.
The manual includes sections on sanitary mounting, cable termination (standard and CE-specific), troubleshooting, and warranty. It emphasizes proper cable construction to avoid electromagnetic interference and provides diagrams for mounting and termination. Calibration and installation procedures are also covered, including alignment of the sensor with product flow direction.
| Wetted Materials | PVDF or PFA Teflon |
|---|---|
| Operating Temperature Range | 14 to 347°F (-10 to 175°C) |
| Measuring Range | 0-200 to 0-2,000,000 microSiemens/cm |
| Temperature Compensator | Pt 1000 RTD |
| Sensor Cable | 5 conductor (plus two isolated shields) cable with Teflon-coated jacket; rated to 175°C (347°F); 20 ft. (6 m) long |
| Pressure/Temperature Limits | 200 psi at 347°F (13.8 bar at 175°C) |
The sensor is constructed of PVDF or PFA Teflon depending on option choice.
The sensor measures conductivity from 0-200 to 0-2,000,000 microSiemens/cm.
The operating temperature range is 14 to 347°F (-10 to 175°C).
The sensor may be installed in a special 2" x 2" sanitary tee, or in standard 2 1/2" x 2", or 3" x 2" short outlet reducing tee.