This manual provides instructions for installing, connecting, configuring, and using the L3 pressure and level transmitter.
This manual provides detailed instructions for installing, wiring, configuring, and maintaining the L3 pressure and level transmitter. It includes specifications, warnings, dimensional drawings, and step-by-step procedures for sensor and instrument wiring, tank configuration, alarm settings, and calibration. The device is designed for liquid measurement in food and beverage applications, with features like IP67/NEMA 4X protection and Hart 7.0 compatibility.
| Measuring Range (bar) | 0...0.4 / -1...2 / -1...7 / -1...35 |
|---|---|
| Overpressure Strength | 1.5 x nominal pressure of measuring element |
| Measurement Accuracy | ≤ 0.10 % in calibrated measuring range (turndown to 5:1) |
| Temperature Range (Process) | -18...110 °C (0...230 °F) |
| Response Time | < 0.1 seconds |
| Output Signal | 4...20mA and Hart 7.0 |
| Protection Class | IP 67 (cable gland) / NEMA 4X (M12 plug) |
| Auxiliary Voltage | 18...35 V DC |
| Tightening Torque | 27 Nm (20 ft-lbs) |
| Weight | approx. 780 g |
The overpressure strength is 1.5 times the nominal pressure of the measuring element.
The response time is less than 0.1 seconds.
It supports 4...20mA analog output and Hart 7.0 communication.