This document is the instruction manual of KEW 3021 3022 3023 insulation - continuity tester
This manual provides operating instructions for the KEW 3021, 3022, and 3023 insulation and continuity testers. It covers safety warnings, measurement procedures, instrument layout, and maintenance. The testers are designed for low-voltage insulation testing up to 600V and include continuity check with audible feedback. Safety precautions are emphasized, including CAT III 600V compliance and warnings against wet conditions or exposed metal. The manual also details specifications, range settings, battery and fuse replacement, and cleaning procedures.
The device features auto-ranging, backlight, memory storage for up to 99 measurements, and auto-power-off to conserve battery life. It includes a safety key for selecting 500V or 1000V ranges and a discharge function for capacitive loads. The manual includes detailed specifications for each model’s voltage and resistance ranges, accuracy, and environmental tolerances.
| Operating error | ±30% (per IEC61557) |
|---|---|
| Voltage range | 20~600V |
| Accuracy (Voltage) | ±(3%rdg + 6dgt) |
| Auto-discharge function | Yes |
| Continuity check | Yes, with audible buzzer |
| Battery life | Auto-power-off after 15 minutes |
| Safety standard | IEC 61010-1 (CAT III 600V) |
The tester can measure voltages up to 600V AC/DC.
Yes, it can store up to 99 measurements in internal memory at the insulation resistance range.
The instrument automatically turns off after approximately 15 minutes of inactivity.