Triatek HMS-1650 Fume Hood Controller Installation Manual

This file is the HMS-1650 Fume Hood Controller User Manual, which introduces the features and usage of the product.


Triatek HMS-1650 Fume Hood Controller Installation Manual - cover page
Brand
TRIATEK
Category
Industrial Supply
Document type
Installation Guide
Language
English
Pages
48
File format
PDF
File size
7.8 MB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
F4E209D49610091A269BA8BB6027D315

About this manual

This document is the installation manual for the Triatek HMS-1650 Fume Hood Controller. It provides instructions for initial configuration, including network settings, sensor input ranges, and analog output configurations for controlling exhaust dampers. The guide details the physical installation process, including mounting dimensions and wiring diagrams for various actuator types.

What's inside

  • CONTROLLER SETTINGS
  • Wiring Diagrams
  • Mounting and Installation
  • Sensor Types and Mounting
  • Communications (BACnet/Metasys)
  • BASIC PROGRAMMING
  • Module Settings
  • Cleaning the Display
  • Flow Diagrams

Specifications

Analog Inputs4-20mAdc, 0-5Vdc or 0-10Vdc
Analog Outputs4-20mAdc, 0-5Vdc or 0-10Vdc
Relay Outputs1A@24VDC
Power SupplyClass 2, 24Vac ±10%, 30VA universal 120/240 to 24 Vac
LCD Resolution240 x 320 portrait
Operating Temperature32° to 125° F
Face Velocity Accuracy0-200 FPM ± 2 FPM*

Frequently asked questions

What is the minimum wire size for the control signal?

The minimum wire size required for the control signal is 18 AWG.

What transformers are required for the controller and actuator?

A single 120/24V 30Va transformer is required for the controller, and a separate 120/24V 20Va transformer is required for the actuator.

Can the controller operate in silent mode?

Yes, the audible alert can be configured to operate in silent mode.

Download PDF (7.8 MB)