《Halogen Calibrated Light Sources Installation and Operation Manual》details the features, installation, and operation of Ocean Optics Halogen Calibrated Light Sources series products
This manual provides installation and operation instructions for the Ocean Optics HL-3 Series Halogen Calibrated Light Sources. It covers unpacking procedures, connecting the light source to spectrometers using either a cosine corrector or an integrating sphere, and details on bulb replacement limitations. The document also includes a use log template for tracking lamp usage.
| Power Input | 12 VDC/800 mA – 1100 mA |
|---|---|
| Bulb Life | 10,000 hours |
| Time to Stabilized Output | 15 minutes |
No, you cannot change the bulb in this light source as a recalibration is required when the bulb is replaced.
The light source must be recalibrated after 50 hours of use.
The maximum uncertainty level for the HL-3 plus models is 3%.