The DD2011 dental separator is disposable and is manufactured by SimpleOne. The separator has passed the ISO 11143 protocol certification and is suitable for type 3 amalgam separator. The separator has a maximum water flow rate of 1.0 liters per minute and a maximum fillable volume of 360 ml. The separator is installed in the dental vacuum line, and a suction loss will occur 3-7 days before the DD2011 becomes full and requires replacement. The DD2011 needs to be changed when the air flow at the HVE tip is no longer adequate, which will be demonstrated by the noticeable decrease of suction. The decrease in suction is the gauge used to determine the necessary replacement of the unit. If the system is not replaced within 3-7 days after noticeable decrease of suction is noted and is allowed to fill completely, the system's unique design will alarm you by shutting off completely, not allowing hazardous waste to flow past the unit. The change out rate varies from 9 to 12 months and depends upon chair activity. The 2011 is designed to handle a standard dental vacuum of 7 inches of Hg. Instructions for changing of the amalgam separator are found at the end of the manual.
This manual provides instructions for installing and maintaining the Dental EZ DD2011 chairside amalgam separator. It details how to connect the unit to the dental vacuum line, replace the separator when suction decreases, and handle hazardous waste. The document also includes guidance on avoiding premature failure, such as avoiding foam-producing cleaners and 90-degree fittings. Safety precautions for handling mercury-containing waste are emphasized.
The DD2011 is a Type 3 amalgam separator designed for dental chairs with a standard vacuum of 7 inches of Hg. It has a 360 ml fillable volume and a maximum water flow rate of 1.0 liter per minute. The separator shuts off completely when full to prevent hazardous waste from passing through. Replacement intervals vary from 9 to 12 months depending on chair activity.
| Type | Type 3 amalgam separator |
|---|---|
| Maximum Water Flow Rate | 1.0 liters per minute |
| Fillable Volume | 360 ml |
| Designed Vacuum | 7 inches of Hg |
Replace it when the air flow at the HVE tip is no longer adequate, which is indicated by a noticeable decrease in suction. The unit will shut off completely if allowed to fill fully.
The dental chair should have a vacuum level of 21 to 27 kPa (6 to 8 in-Hg) with a minimum of 21 kPa (6 in-Hg) at the farthest HVE inlet.