GSI 39 Auto Tymp User Guide User Guide
The GSI 39 Auto Tymp is a medical device designed for compliance with IEC and UL 60601-1 standards when used in the patient vicinity. It functions as both an audiometer and a tympanometer, requiring the use of properly grounded plugs and receptacles for patient and operator safety. The device is specifically calibrated and should only be serviced by authorized Cardinal Health/GSI technicians.
| Standards | UL 60601-1, IEC/EN 60601-1, CSA C22.2 No.601-1-M90 |
|---|---|
| Probe Tone (226 Hz) | 226 Hz, ±2% |
| Probe Tone (1000 Hz) | 1000 Hz ±2% |
| Sound Pressure Level (226 Hz) | 85.5 dB SPL, ±2.0 dB |
| Sound Pressure Level (1000 Hz) | 75 dB SPL, ±2.0 dB |
| Pneumatic System Pressure Range | +200 to -400 daPa |
Service adjustments should be done only by an authorized Cardinal Health/GSI service technician.
The system is intended for continuous operation and has a protective classification of Class 1 Type B.
The compliance accuracy is ±0.1 cm3 or ±5%, whichever is greater.