Conductance Tester User Guide
The YUASA GYT117 conductance tester is designed to screen the state-of-health for Valve Regulated Lead Acid (VRLA) batteries ranging from 0.5Ah to 25Ah. It measures and displays DC voltage and conductance in Siemens (S), with voltage test limits set for 6V and 12V batteries. The device includes safety precautions and step-by-step instructions for testing, including how to select voltage, set reference values, and interpret results. Temperature compensation scales are provided to adjust readings based on battery temperature.
| Voltmeter operating range | +6.0 to +19.99 Vdc |
|---|---|
| Voltage accuracy | +/–50 mV across operating range |
| Operating temperature | –18 to 50°C (0 to 120°F) |
| Voltage test limits (12V high) | 13.80V |
| Voltage test limits (12V low) | 12.00V |
| Voltage test limits (6V high) | 6.90V |
| Voltage test limits (6V low) | 6.00V |
| Siemens range | 20 to 1200S |
–18 to 50°C (0 to 120°F).
Press the DOWN ARROW button to select 6V, then press ENTER to start the test.
The battery has exceeded the tester’s operating range. Refer to the specifications for limits.
Multiply the reference value by the appropriate compensation factor based on temperature (e.g., 0.930 at 35°C, 1.175 at 0°C).