Vaisala HMP45A/D Humidity and Temperature Probes User's Guide

This document is the user manual of Vaisala HMP45A/D humidity and temperature probes, introducing the probe's usage methods, functions and features


Vaisala HMP45A/D Humidity and Temperature Probes User's Guide - cover page
Brand
Vaisala
Category
Other Electronics
Document type
User Manual
Language
English
Pages
10
File format
PDF
File size
417 KB
Published
18 October, 2023
Updated
18 October, 2023
Model
HMP45A
MD5 checksum
5163945ADC27CB1BB79E1B271BEE0D17

About this manual

The Vaisala HMP45A and HMP45D probes are designed for measuring relative humidity and temperature. The HMP45A features an active voltage output for temperature, while the HMP45D uses a passive resistive output (Pt 100). Both models utilize a HUMICAP®180 capacitive sensor for humidity and platinum sensors for temperature, protected by a standard membrane filter.

What's inside

  • General
  • Connections
  • Calibration and Maintenance
  • Humidity Calibration
  • Changing the Humidity Sensor
  • Spare Parts and Accessories
  • Technical Data
  • Electromagnetic Compatibility
  • Guarantee

Specifications

Humidity Measurement Range0.8...100 %RH
Humidity Accuracy at +20 °C+1 %RH (factory)
Response Time (90%)15 s

Frequently asked questions

How often should the probes be calibrated?

Calibration and maintenance should be performed at regular intervals, with a recommended interval of one year.

Can the HMP45D be used with the HMI41 indicator?

No, the HMP45D cannot be used with the HMI41 indicator because the passive temperature signal results in error messages.

What is the maximum cable length for differential measurement?

With SIGNAL GROUND, the cable can be extended up to 100 metres without disturbing measurement accuracy.