About this manual

The Geotech Interface Probe with Float is a portable instrument designed to accurately measure water and hydrocarbon layer thickness in monitoring wells. The unit consists of a stainless steel and PTFE probe attached to a reel-mounted, HDPE coated engineer's tape. It features an audible signal and a visible red LED that activate when the probe reaches water or product, with a green LED and oscillating tone for conductive fluids like water. The device includes an auto shutoff circuit to prevent battery discharge after 10 minutes of continuous operation.

What's inside

  • System Description
  • System Installation
  • System Operation
  • System Maintenance
  • System Troubleshooting
  • System Specifications
  • System Schematic
  • Replacement Parts List

Specifications

Tape Accuracy1/100 of a foot per 100'
Probe Weight19.75 oz (560 g)
Probe Diameter1.5" (3.8 cm)
Minimum Conductivity Threshold> 6.7µS
Self shut-off time10 minutes
Operating temperature range32 – 140 °F (0 – 60 °C)

Frequently asked questions

What indicates water or conductive fluid detection?

The LED turns green and the tone oscillates.

How long does the auto shutoff circuit last?

The unit shuts off after 10 minutes of continuous operation.

What should be done if the unit is stored for longer than three months?

Remove the battery to prevent battery leakage.