Fluke 53IIB/54IIB Thermometer Users Manual

Fluke Corporation warranty policy


Fluke 53IIB/54IIB Thermometer Users Manual - cover page
Brand
Fluke
Category
Other Electronics
Document type
User Manual
Language
English
Pages
25
File format
PDF
File size
449 KB
Published
18 October, 2023
Updated
18 October, 2023
Models
53IIB, 54IIB
MD5 checksum
59BB0FCF26E52B42A1505FB5B1472BDC

About this manual

The Fluke 53IIB and 54IIB are microprocessor-based digital thermometers designed to use external J-, K-, T-, E-, R-, S-, and N-type thermocouples. The device features a display for primary and secondary readings, including minimum, maximum, and average values, as well as memory storage for logged data. Users can configure setup options such as logging intervals, thermocouple types, offsets, and sleep modes via dedicated buttons.

What's inside

  • Safety Information
  • Getting Started
  • Components
  • Display Elements
  • Using the Thermometer
  • Changing Setup Options
  • Measuring Temperatures
  • Communicating with a PC
  • Maintenance
  • Specifications

Specifications

Supported Thermocouple TypesJ-, K-, T-, E-, R-, S-, and N-type
Operating Temperature-10 o C to 50 o C
Storage Temperature-40 o C to +60 o C
Battery Requirement3 AA batteries
Weight280 g (10 oz)
Dimensions2.8 cm × 7.8 cm × 16.2 cm
Measurement Accuracy (J/K/T/E/N)±[0.05 % of reading + 0.3 o C]
Measurement Accuracy (R/S)±[0.05 % of reading + 0.4 o C]

Frequently asked questions

How often should the thermometer be calibrated?

Fluke recommends that you calibrate the thermometer annually, starting one year after purchase.

What is the maximum differential common mode voltage?

The maximum voltage difference between T1 and T2 is 1 V.

Can the thermometer communicate with a PC?

Yes, you can transfer memory contents to a PC using FlukeView Forms via an IR serial connection.