Fresenius Medical Care 2008T Hemodialysis Machine bibag System Operator's Instructions

This document is about the operator's instructions for 2008T hemodialysis machine bibag system.


Fresenius Medical Care 2008T Hemodialysis Machine bibag System Operator's Instructions - cover page
Brand
Fresenius Medical Care
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
927 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
F1125C4ED44C518297B3413E0A49C276

About this manual

The bibag system is a hardware option for the Fresenius Medical Care 2008T hemodialysis machine that uses a bibag disposable filled with dry bicarbonate powder to generate dialysate solution. The disposable connects to the machine through a connector to the left of the concentrate ports, and the “45x” concentrate family must be selected in Service Mode to use it.

The operator’s instructions cover general warnings, using the bibag system for a dialysis treatment, changing the disposable during treatment, emptying it at the end of treatment, using the blue bicarbonate connector, cleaning the bibag connector, and 2008T machine with bibag system specifications.

What's inside

  • General Warnings
  • Selecting concentrates in Service Mode
  • bibag System Details
  • Using the bibag system for a dialysis treatment
  • Changing the bibag disposable during treatment
  • Emptying a bibag disposable at the end of treatment
  • Using the blue bicarbonate connector
  • Cleaning the bibag connector
  • 2008T Machine with bibag System Specifications

Specifications

Required concentrate family45x
Functional Board software2.35 or greater
Actuator Board software2.15 or greater
Maximum acid and bicarbonate pressure with Central Delivery System10 psi
Required filterDiasafe Plus filter
Final dialysate pH range6.9-7.6
Final dialysate conductivity tolerancewithin +5% of the nominal machine value

Frequently asked questions

What concentrate family must be selected to use the bibag system?

The “45x” concentrate family must be selected in Service Mode.

Which filter is required when the bibag system is in use?

The Diasafe Plus filter is required when the bibag system is in use.

What is the maximum acid and bicarbonate pressure when using a Central Delivery System?

The acid and bicarbonate pressures must not exceed 10 psi.

Where does the bibag disposable connect on the 2008T machine?

It connects to a special connector to the left of the concentrate ports.