This manual introduces the Neonatal NAVA Management Protocol, including the selection, placement and positioning of the catheter.
This document outlines the procedure for placing and positioning the Edi catheter for Neonatal NAVA (Neurally Adjusted Ventilatory Assist) management. It details the specific steps for connecting the Edi module, selecting the correct catheter size based on patient weight and height, and verifying catheter position using waveform analysis and color indicators. The guide also specifies the required ventilator settings for both invasive and non-invasive NAVA modes on the Servo-i ventilator.
| NAVA Level | 2 cmH2O/mcV |
|---|---|
| PEEP | 5-6 cmH2O |
| Trigg Edi | 0.5 mcV |
| Backup PC | above PEEP |
| Backup Resp Rate | 40-60 breaths/min |
| Apnea Time | 2 sec |
| Pressure Alarm Lower | 35-40cmH2O |
| Minute Volume Lower | 0.01 l/min |
| Minute Volume Upper | 5.0 l/min |
| Respiratory Rate Lower | 15 b/min |
| Respiratory Rate Upper | 100 b/min |
| End Expiratory Pressure Lower | 2 cmH2O |
| End Expiratory Pressure Upper | 10 cmH2O |
Select the catheter size based on the patient's weight and height using the provided chart (e.g., 6 Fr/49cm for 500g—1.5kg).
Proper positioning is indicated by a diminishing ECG waveform progressing from the 1st to the 4th waveform and the presence of a pink color in the 2nd and 3rd waveforms.